Product Details
Place of Origin: Shanghai,China
Brand Name: Macy
Certification: CE.Rosh, ISO
Model Number: UV-1800APC
Payment & Shipping Terms
Minimum Order Quantity: 1 pc
Packaging Details: One carton of 38KGS
Delivery Time: In one week
Payment Terms: T/T, L/C, Western Union
Supply Ability: 1 unit one day
Band Width: |
1nm |
Wavelength Range: |
190-1100nm |
Wavelength Accuracy: |
±0.3nm |
Stray Light: |
≤0.03% τ (220nm NaI,340nm NaNO2) |
Noise: |
± 0.0002A @500nm |
Dispaly: |
6 Inches High Light Blue LCD |
Band Width: |
1nm |
Wavelength Range: |
190-1100nm |
Wavelength Accuracy: |
±0.3nm |
Stray Light: |
≤0.03% τ (220nm NaI,340nm NaNO2) |
Noise: |
± 0.0002A @500nm |
Dispaly: |
6 Inches High Light Blue LCD |
●UV-1800APC successful implementation of the stringent requirements of high accuracy and reliability measurement to meet the requirements of various applications that can be used in biological research, bio-industry , pharmaceutical analysis , pharmaceutical, teaching and research, environmental protection , food hygiene, clinical examination , health and epidemic prevention , and other fields .
●A wide wavelength range, the wavelength range to meet the various requirements of the field
●Automatic design to achieve the most simple means of measurement
●LSI design greatly improves the scalability and reliability of the system
●Improved optimization of the optical design , import source and receiver system created a high performance and high reliability
●Rich measurement methods , with a wavelength scan , time scan , multi-wavelength determination , multi-order derivative determination (optional ) , dual-wavelength , three-wavelength (optional ) DNA protein measurements (optional ) , and other measurement methods to meet different measurement requirements , and can be displayed directly on a 6- inch screen
●Based on user requirements optional hole rack , manual four with stand , manual Eighth rack , automatic Eighth frame, glass stand , test tube rack , 1cm colorimetric frame , 5cm colorimetric frame , 10cm colorimetric racks
●Measurement data can be output through the printer with USB interface
●Can be powered down to save the measurement parameters and data , user-friendly
●Can achieve more accurate and flexible measurement via PC control , which can meet the needs of different users
Technical Specifications
Optical System | Double beam |
Wavelength Range | 190~1100nm |
Band Width | 1nm |
Wavelength Accuracy | ±0.3nm |
Wavelength Repeatability | ≤0.1nm |
Photometric Accuracy | ±0.3% τ(0-100%τ) ±0.002A(0~0.5A) ±0.003A(0.5A~1A) |
Photometric Repeatability | 0.15% τ(0-100%τ) ±0.001A(0~0.5A) ±0.0015A(0.5A~1A) |
Stray Light | ≤0.03% τ (220nm NaI,340nm NaNO2) |
Stability | 0.0005A/h@500nm |
Noise | ± 0.0002A @500nm |
Baseline Flatness | ± 0.001A |
Photometric Mode | T,A,C,E |
Wavelength Setting | Automatic |
Photometric Display Range | -4~4A |
Display Mode | 6 inches high brightness blue LCD |
Detector | Import Silicon Photodiode |
Light Source | Import Deuterium Lamp & Tungsten Halogen Lamp |
Power Requirement | AC 220V/50Hz 110V/60Hz |
Power | 120W |
Dimensions(W*D*H) | 560×450×230mm |
Weight | 28Kg |